Purity oxygen analyzer

Purity oxygen analyzer uses an imported limiting current sensor with a rapid response speed, high accuracy and long calibration cycle for features. The frame mount installation mode makes it simple and convenient to install.Purity oxygen analyzer main features:1.No reference gas is needed
Products Inquiry

Product Details

Purity oxygen analyzer uses an imported limiting current sensor with a rapid response speed, high accuracy and long calibration cycle for features. The frame mount installation mode makes it simple and convenient to install. 
Purity oxygen analyzer main features:
1.No reference gas is needed and analyzer will not be influenced by oxygen purity in atmosphere;
2.With the functions of compensation and sensor protection built in, the analyzer can be on power under 99.99% occasion without affecting the sensor’s serving life.
3.With the function of temperature compensation, the amount of sample gas flow wouldn’t bring any error to the sensor’s measured results;
4.Altitude measurement module can be expanded to avoid the influence from atmospheric pressure change on the measured accuracy;
5.The measured accuracy of the whole range can be met by a random point calibrated, making the operation simple and convenient.


Purity oxygen analyzer technical parameters:
 Test principle ion current (liming current )
 Display mode 4 bits LED display
 Test range 10.00% ~ 99.99% O2
 Test range 0.01%
 Accuracy ≤±1%FS
 Repeatability ≤±1%
 Response time T90≤30S
 Analog output

4~20mA.DC(non-isolated output, loaded resistance lower than 500 Ohm)

 0~5V.DC(non-isolated output, loaded resistance higher than 10k Ohm)

1 way programmable alarm output (Dry contact type, maximum load 220VAC/2A)

 Communication port RS232(default) or RS485
 Power supply AC85 ~ 264V 50/60Hz
 Sampling mode sample gas pass-in
 Sample gas temperature -10 ~ +50℃
 Sample gas flow 300 ~ 600ml/min
 Sample gas pressure 0.05Mpa≤entrance displayed pressure≤0.35Mpa,steady pressure environment
 Background gas Nand inactive gas
 Appearance Size 110mm×190mm×235mm(H×W×D)
 Install Hole Cut 90mm×170mm (H×W)
 Serving life type C>5 years, type D>3 years(under   normal use)
 Gas circuit port Φ6 plastic pipe junction  (sensor built-in{)
 Install mode frame mount
Disclaimer: The product introduction content (including product images, product descriptions, technical parameters, etc.) of this site is for reference only. As the update may not be timely, there will be some differences between the content and the actual situation. Please contact our sales staff to confirm. The information provided on this site does not constitute any offer or promise, and the company will improve and modify any information on the website from time to time without prior notice.